Sampling Interval |
N/A ~4 Seconds |
Selectable Alarm Delay Periods |
N/A 0 minute1/2 minute1 minute5 minute10 minute30 minute |
Selectable Audible Alarm Delay Periods (Mute or Disable) |
N/A 10 minute15 minute20 minute30 minute5 minute60 minute |
Output Relay (Dry Contacts) |
N/A Single Form C (Single Pole Double Throw (SPDT)),Normally Open (NO) or Normally Closed (NC), user adjustable delay period |
Battery Backup |
N/A 2 ea. AA 1.2V NiMH 2500 mA·h, 30 mA built-in trickle charger, ~12-hour life |
Dry Contact Inputs |
N/A Normally Closed (NC) or Normally Opened (NO) |
Alternating Current (AC) Input Voltage |
N/A 100 to 240 V |
Input Frequency |
N/A 50 to 60 Hz |
Root Mean Square (RMS) Current at 115 Volt (V) Alternating Current (AC) Voltage |
N/A < 0.35 A |
Note for Power Adapter Input |
N/A UL listed, 2-prong US |
Direct Current (DC) Output Voltage |
N/A 12 V |
Output Current |
N/A 0.6 A |
Output Power |
N/A 7.2 W |
|
N/A
|
|
N/A
|
|
N/A
Eliminate nuisance and transient condition alarms with the ADM-215 Remote Alarm Delay Module. The ADM-215 allows a technician to set a time that must elapse, with the alarm condition continuously present, before the alarm signal is relayed to a central monitoring system or dialer. By selecting the optimum alarm delay, transitory fluctuations in power, temperature (for example, due to loading or unloading a ULT freezer), and other monitored parameters are filtered out and nuisance alarms are eliminated. |